Baxter AS50 Features:
The Auto Syringe AS50 Infusion Pump is designed to meet the fluid and drug delivery requirements of today's changing clinical environment. It provides an accurate, continuous or intermittent infusion of intravenous solutions, drug solutions, whole blood, and packed red blood cells. The pump is indicated for infusion via intravenous (IV), intra-arterial (IA), epidural, or subcutaneous routes of administration. It can be piggybacked into an ongoing infusion line to deliver secondary solutions automatically, precisely, and economically. Infusion rates are programmable from 0.01 to 438 mL/hr.
The pump accepts standard disposable syringes from 1 mL to 60 mL in size. A numeric keypad simplifies programming and makes the pump easy to use. During programming, the status display prompts the user to enter data about the syringe and the infusion. While the pump is running, the status display shows information about the infusion. Safety and effectiveness are reinforced by pre-programmable bolus operation, titration of a dose without interruption of fluid flow, and easily understood alarm and alert messages. The pump includes an optional serial communication feature, which provides remote computer control and monitoring capabilities.
- Digital syringe pump.
- Provides accurate infusion of intravenous solutions, drug solutions, whole blood, and packed blood cells.
- Indicated for the following routes of administration: intravenous, intra-arterial, epidural or subcutaneous.
- Accepts standard disposable syringes from 1mL to 60 mL.
Specifications
Baxter AS50 Syringe Pump Specifications:
Dimension
- Height: 3.4 in ( 8.6 cm)
- Width: 2.6 in (6.7 cm)
- Length: 10 in (25 cm)
- Weight: 3.2 lbs. (1.45 kg)
Flow Rate Accuracy
- ± 3% of full scale plunger travel. (not including syringe tolerance.)
Volume Accuracy
- ± 3%, or 0.007" of travel, whichever is greater. (not including syringe tolerance.)
Syringes
- B-D: 1, 3, 5, 10, 20, 30, 60 mL plastic
- Sherwood Monoject: 1, 3, 6, 12, 20, 35, 60 mL
- Terumo: 1, 3, 5, 10, 20, 30, 60 mL
Flow Rate Range
- 0.01 mL/hr to 438 mL/hr, depending on the syringe.
Delivery Volume
- Full syringe volume for 1-60 mL syringes.
Data Display
- Self-prompting, multi-field.
Status Display
- Nine-LED array.
Power Requirements
- AC: 105-125V 60 Hz (battery charger)
- DC: internal rechargeable battery pack.
Battery Operating Time
- 5 hours of operation at 100 mL/hr.
- 12 hours of operation at 2 mL/hr using a 60 mL syringe, following a charge of not less than 16 hours.