Casmed CAS 740 Vital Signs Monitor w/ FlexiPort Reusable Blood Pressure Cuff 10" 11" 12 Features
The CasMed 740 monitor is a multi-parameter monitor measuring blood pressure, oxygen saturation, and temperature. Non-invasive blood pressure is calculated using the oscillometric technique determining systolic, diastolic, mean arterial pressure, and pulse rate. The pulse oximeter function continuously monitors and displays values for functional arterial hemoglobin saturation and pulse rate. Temperature is obtained in the normal (predictive) mode in as little as four (4) seconds. A monitoring mode is available for taking axillary temperatures.
This monitor is a portable device intended to be used by trained clinicians for multi-parameter vital signs monitoring of neonatal, pediatric, and adult patients in health care bedside applications as well as for intra-facility or inter-facility and EMS transport. The parameters displayed are the non-invasive blood pressure (systolic, diastolic, and mean arterial pressure), pulse rate, functional oxygen saturation of arterial hemoglobin (%SpO2), and temperature.
It has been designed to promote patient safety. The maximum amount of time allowed to complete a blood pressure measurement is 120 seconds in adult mode and 90 seconds in neonate mode. If the measurement has not been completed within that time, the cuff is deflated automatically, and a message is displayed indicating the problem.
Specifications
NIBP MEASUREMENT
Technique: Oscillometric (MAX NIBP™ Technology)
Microprocessor software eliminates most ambient noise and motion artifacts.
Patient Range: Neonate – Adult
Blood Pressure Range NEO ADULT
Systolic: 30 – 135 mmHg 30 – 255 mmHg
Diastolic: 15 – 110 mmHg 15 – 220 mmHg
MAP: 20 – 125 mmHg 20 – 235 mmHg
Pulse Rate Range: 40 – 240 BPM 30 – 240 BPM
Accuracy
Blood Pressure: +/-5 mmHg with a standard deviation no greater than 8 mmHg (See Standards)
Pulse Rate: +/-2% or +/-2 BPM, whichever is greater
OXIMETRY (OPTIONS)
Masimo SET®
Type: Functional Oxygen Saturation
SpO2 % Range: 0 - 100%
SpO2 Accuracy: 70 - 100%, +/-2 digits (1 S.D.) Adult
70 - 100%, +/-3 digits (1 S.D.) Neonate
Measurement
Wavelengths: Red 660 Nanometers
Infrared 905 Nanometers
Power: Maximum radiant power at 50 mA pulsed is 0.79mW
Pulse Rate Range: 25 - 240 BPM Pulse Rate Accuracy: +/-3 BPM Numerics: Updated every 50 milliseconds max
Adult and Infant Cuffs
- Infant - 6 cm (2.4 in.) 11 - 19 cm (4.3 - 7.5 in.)
- Small Child - 7 cm (2.8 in.) 15 - 24 cm (6.0 - 9.4 in.)
- Child - 9 cm (3.5 in.) 18 - 28 cm (7.1 - 11.0 in.)
- Small Adult - 12 cm (4.7 in.) 18 - 28 cm (7.1 - 11.0 in.)
- Adult - 14 cm (5.5 in.) 24 - 37 cm (9.4 - 14.6 in.)
- Large Adult - 16 cm (6.3 in.) 26 - 42 cm (10.2 - 16.5 in.)
POWER
Source: External line or internal battery
AC Power: 100 - 240 VAC, 50/60 Hz, 0.5A; Fuse Rating – T500mAL250V (two provided
DC Power (EMS Option): +12 VDC; 7W; Fuse Rating – Wickman Type TE5, Time Lag, 3.15A, 125VAC or approved equivalent (one provided)
Battery: Smart Nickel Metal Hydride (NiMH) battery pack (user removable) Charge Time: 4 hours Operation on battery: 100 NIBP readings when set in the 5-minute Automatic Mode
Leakage Current: 100 microamps (maximum)
TEMPERATURE (OPTIONAL)
Temperature Range: 28.9 to 42.2 °C (84.0 to 108.0 °F) Accuracy: +/-0.1°C (+/-0.2°F), Meets or exceeds ASTM Standards
Monitors may not meet performance specifications if stored or used outside temperature and humidity ranges. When moving the monitor from a storage location, wait at least one hour prior to use to allow the monitor to adjust to room temperature.
PHYSICAL DIMENSIONS & WEIGHT
Base Unit H x W x D: 6.75 in x 8.5 in x 3.0 in
(17 cm x 21.5 cm x 7.5 cm)
Weight: 3 lbs approx. (1.4 kg)
SAFETY LIMITS
Automatic Cuff Deflation: If cuff pressure exceeds 290 mmHg (Adult); 145 mmHg (Neonate) If measurement time exceeds 120 seconds (Adult), 90 seconds (Neonate) If the safety timer detects microprocessor failure
BATTERY POWER