BioMedical-Cart Cart
0 item(s) items
Phone +1-800-910-1422
Hours of Operation: Mon. – Sat. 8:00 a.m. – 6:00 p.m. PDT
A Biomedical Service Covidien Kangaroo 924 Enteral Feeding Pump
A Biomedical Service Covidien Kangaroo 924 Enteral Feeding Pump
A Biomedical Service Covidien Kangaroo 924 Enteral Feeding Pump
A Biomedical Service Covidien Kangaroo 924 Enteral Feeding Pump
A Biomedical Service Covidien Kangaroo 924 Enteral Feeding Pump

Covidien Kangaroo 924 Enteral Feeding Pump

SKU: 120C-60HZ1AMP, 392455
3 Months Limited Warranty
Certified Used
Regular price
Price
$220.00
Unit price
per 
iso
Buy It Now

Covidien Kangaroo 924 Enteral Feeding Pump features:

The Covidien Kangaroo™ 924 Enteral Feeding Pump with Pole Clamp is a rotary peristaltic pump designed to regulate the flow rate of enteral feedings. Among the features provided by this pump are:

• Portable operation-built in rechargeable battery

• Wide flow rate range 1-300 mL/hr in 1 mL/hr increments

• Small and lightweight for easy transport

• Language Selection (13 languages available)

• Touch panel and knob operation

• Shut off and activation of audible and visible alarms when:

− Feeding container is empty − Feeding tube becomes occluded

− Drip sensors blocked

− Battery low (alarms 15 minutes before pump shut off)

− Unit is left on “Hold” longer than 5 minutes

− Feeding set is improperly loaded − Self-diagnosed problem which requires technical servicing

− Pump motor malfunction

− Ambient light error

• Automatic systems check after the pump has been turned ON

• Large LED display

• Low occlusion pressure. The pump will not overcome back pressure greater than approximately 15 psi (103.4 kPa)

• Integral pole clamp

• Use of Optix feeding sets to prevent free flow from inadvertent misload of the feeding set onto the pump

• Pump running indicator lights

• DC motor with intermittent operation

• Adjustable alarm volume

• Memory of flow rate setting is retained for 16 or 24 hrs. after the pump shut off, depending on AC connection status.

Specifications

Type Infusion Device

Volumetric

Pumping Mechanism

Rotary Peristaltic

Feeding Set Required

Covidien Kangaroo 924 Enteral Feeding Pump with Pole Clamp Feeding Sets (see www.covidien.com for details on compatible administration sets)

Operating Range Rate:

1-300 mL/hr in 1 mL/hr increments Dose: 1-2000 mL in 1 mL increments Volume Delivered: 1-9999 mL (rollover is accounted for)

Accuracy

10% or 0.5 mL/hr, whichever is larger, with the top of the fluid column at 71.12 cm (28”) ± 0.76 cm (0.3”) above the top of the pump, at a room temperature of 22 °C ± 2 °C (72 °F ± 3 °F), using a new 924 feeding set for no longer than the recommended 24 hours of maximum usage. Confidence limits for accuracy are based upon those included in the ANSI/AAMI ID26-1992 American National Standard for Infusion Devices. Reported accuracy is based upon the time-based accuracy calculation in this standard, as applicable for enteral feeding

Occlusion Pressure

15 psi (103.4 kPa) Nominal

Alarms

• Flow Error

• Low Battery

• Hold Error

• Load Set

• System Error

• VTBD Delivered

The memory of Settings When Pump is Off

16 hours for battery, 24 hours for AC

Battery

• Rechargeable NiMH

• Automatically charges when the pump is plugged into a wall outlet. The recharge time for a fully discharged battery is approximately 12 hours.

• A new battery, when fully charged, will operate for approximately 16 hours at 125 milliliters per hour.

Dimensions Approximately

7.25” High x 6.61” Wide x 4.33” Deep (18.4 cm High x 16.8 cm Wide x 11 cm Deep)

Parts

Similar Items